Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8025B: (1 year) CAL VER /5700,5725 DATE: 03-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 147 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # T.O. 33A1-12-1300-1 # Fluke 8025B instruction manual (PN 824482 Feb. 1988). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R J F W 1.002 ASK+ K 1.003 MMFC C2 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DISPLAY TEST} 1.007 DISP Turn the UUT on. 1.007 DISP AS the UUT is turned ON, verify that all display 1.007 DISP segments momentarily light up. 1.008 EVAL Did the UUT display test pass? 2.001 DISP Allow the UUT to stabilize at: 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 80%. 2.002 HEAD {AC VOLTAGE TESTS} 2.003 DISP Select the ACV Function. 2.003 DISP [32] Connect the 5700A and the UUT as follows. 2.003 DISP [32] 5700A OUTPUT HI to UUT V Ohm Diode 2.003 DISP [32] 5700A OUTPUT LO to UUT COM 2.003 DISP [32] 5700A SENSE HI to UUT V Ohm Diode 2.003 DISP [32] 5700A SENSE LO to UUT COM 2.003 DISP [32] 5700A AUX CURRENT OUTPUT to UUT mA/uA 2.004 HEAD AC VOLTAGE TESTS: {3.2V Range} 2.005 5700 3.2 2.700V 0.016U 100H 2W 3.001 HEAD AC VOLTAGE TESTS: {32V Range} 3.002 5700 32 27.00V 0.57U 10kH 2W 4.001 HEAD AC VOLTAGE TESTS: {320V Range} 4.002 5700 320 250.0V 5.3U 10kH B1 2W 5.001 HEAD AC VOLTAGE TESTS: {1000V Range} 5.002 5700 1000 1000V 33U 10kH B1 2W 6.001 HEAD {DC VOLTAGE TESTS} 6.002 DISP Select the DCV Function. 6.003 HEAD DC VOLTAGE TESTS: {3.2V Range} 6.004 5700 3.2 2.700V 0.003U 2W 7.001 HEAD DC VOLTAGE TESTS: {32V Range} 7.002 5700 32 27.00V 0.03U 2W 8.001 HEAD DC VOLTAGE TESTS: {320V Range} 8.002 5700 320 250.0V 0.3U 2W 9.001 HEAD DC VOLTAGE TESTS: {1000V Range} 9.002 5700 1000 1000V 2U 2W 10.001 HEAD {RESISTANCE TESTS} 10.002 DISP Select the Ohms Function. 10.003 HEAD RESISTANCE TESTS: {320 Ohm Range} 10.004 5700 320 100.0Z 0.5U CW 11.001 HEAD RESISTANCE TESTS: {3.2 kOhm Range} 11.002 5700 3.2 1.000kZ 0.003U CW 12.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 12.002 5700 32 10.00kZ 0.03U CW 13.001 HEAD RESISTANCE TESTS: {320 kOhm Range} 13.002 5700 320 100.0kZ 0.3U 2W 14.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 14.002 5700 3.2 1.000MZ 0.003U 2W 15.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 15.002 5700 32 10.00MZ 0.11U 2W 16.001 ASK- U 16.002 HEAD {CONDUCTANCE TEST} 16.003 DISP Press RANGE twice to select nanosiemens. 16.004 HEAD CONDUCTANCE TEST: {32nS Range} 16.005 MATH MEM1 = 0 16.006 MEMI Use keyboard to enter actual UUT reading in nS: 16.007 MEME 16.008 MEMC 32 nY 0.10U 17.001 ASK+ U 17.002 HEAD {DIODE TEST} 17.003 DISP Select the DIODE Function. 17.004 EVAL Is the UUT displaying the {overload indicator} (OL)? 18.001 5700 0.090V 0.002U O 2W 19.001 5700 0.090V O S 2W 19.002 EVAL Is the {beeper on}? 20.001 5700 0.110V O S 2W 20.002 EVAL Is the {beeper off}? 21.001 ASK- N 21.002 MESS Slowly decrease the stimulus to 0.6V. 21.002 MESS Verify that the meter produces a short beep as the 21.002 MESS voltage descends through 0.7V (diode threshold). 21.003 5700 1.000V N 2W 21.004 MESS 21.005 EVAL Did the meter produce a short beep through 0.7V? 22.001 HEAD {AC MILLIAMP TESTS} 22.002 DISP Select the AC mA/A Function. 22.003 HEAD AC MILLIAMP TESTS: {32mA Range} 22.004 5700 32 27.00mA 0.42U 50H 2W 23.001 HEAD AC MILLIAMP TESTS: {320mA Range} 23.002 5700 320 270.0mA 4.2U 50H 2W 24.001 HEAD {AC MICROAMP TEST} 24.002 DISP Select the AC uA Function. 24.003 HEAD AC MICROAMP TEST: {3200uA Range} 24.004 5700 3200 2700uA 42U 500H 2W 25.001 HEAD {DC MICROAMP TEST} 25.002 DISP Select the DC uA Function. 25.003 HEAD DC MICROAMP TEST: {3200uA Range} 25.004 5700 3200 2700uA 22U 2W 26.001 HEAD {DC MILLIAMP TESTS} 26.002 DISP Select the DC mA/A Function. 26.003 HEAD DC MILLIAMP TESTS: {32mA Range} 26.004 5700 32 27.00mA 0.22U 2W 27.001 DISP Disconnect the 5700A from the UUT. 27.002 HEAD {DC AMP TEST} 27.003 DISP Connect the 5725A and UUT as follows: 27.003 DISP [32] 5725A CURRENT OUTPUT HI to UUT A 27.003 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM 27.004 HEAD DC AMP TEST: {10A Range} 27.005 5700 10 5.00A 0.06U B1 2W 28.001 HEAD {AC AMP TEST} 28.002 DISP Select the AC mA/A Function. 28.003 HEAD AC AMP TEST: {10A Range} 28.004 5700 10 5.00A 0.09U 1kH B1 2W 29.001 HEAD 29.002 END